DISSOLVED OZONE MONITOR

B&C Dissolved Ozone Monitor provides an economical and reliable measurement system for water quality monitoring and controlling ozone treatment systems.

KEY FEATURES

  • Measuring the display in the range stated by the installed sensor
  • Automatic measurement control function
  • Alarm from the low/high measurement, the setpoint overtime operation, and the logic input;
  • Programmable dual analog output for recording and acquisition of the measurement values or PID regulation;
  • Hold/alarm function activated by two external voltage-free contacts;
  • Automatic /manual auto clean function.

SPECIFICATIONS

Specification Details
Inputs Potentiostatic sensors
Polarographic membrane sensors
Pt100 / Pt1000
Zero ± 20 %, ± 5.0 °C, ± 9 °F
Sensitivity 12.5 ÷ 250 %
Resolution 1 digit
Accuracy 0.2 %
Repeatability 0.1 %
Non-linearity 0.1 %
Dual Filter Software 0.4 ÷ 50.0 seconds for small and large variations
Dual Analog Output 0-20 mA / 4-20 mA Rmax 600 Ω
Dual Set Point HI/LO 0 ÷ 10 %
Delay 0 ÷ 100.0 seconds
Alarm SPDT relay with a delay of 0 ÷ 100.0 seconds
Cleaning Function off / autoclean / manual, relay SPDT
repetition time 0.1 ÷ 100.0 hours
cleaning time 1.0 ÷ 60.0 seconds
holding time 0.0 ÷ 20.0 minutes
SPST and SPDT Relay Contacts 220V – 5 A resistive load
Operating Temperature -10 ÷ 60 °C
Humidity 95% without condensation
Power Supply 85 ÷ 264Vac – 50/60 Hz
9 ÷ 36Vdc, 12 ÷ 24Vac (option 091.42x)
Terminal Blocks Removable
Weight 1360 g
Enclosure ABS, IP 65 protection
Dimensions 256x230x89 mm
EMC/RFI Conformity EN 61326